Technical Specifications: | |
Inductance Measurements : | |
Min : | 0.2 microH |
Max : | 110 H or more |
Ranges : | 8 |
Accuracy : | + 1% for greater ‘Q’ coils |
Capacitance Measurements : | |
Min : | 1 pF |
Max : | 1110 micro Farad or more |
Ranges : | 8 |
Accuracy : | + 1% |
Resistance Measurements: | |
Min : | 100 milli Ohms |
Max : | 10 Megohms |
Ranges : | 8 |
Accuracy : | + 1% |
Display : | Numeric 3.5 digits |
D/QFactor : | Available for Fine adjustment Q: 0-30 D : 0 to 3 |
Excitation : | Internal 1KHz + 10% sine wave and DCPower Supply : 220V +/- 10%, 50Hz. |
Category: Ac bridges
MAXWELL INDUCTANCE BRIDGE – GUPTA BRAND 200655
Optional Accessories :
1) Fixed Frequency Sine Wave Oscillator
2) Sensitive Head Phone /Null Detector
3) Single Dial of 0.1mH
4) Air Core Inductance in Wooden Box Any value from 0.1mH to 100mH
NULL DETECTOR DIGITAL – GUPTA BRAND 200665
The null detector comprises of a solid state amplifier having output stage with meter indication. The said detector has been designed for use with AC bridges replacement of head phone
OWEN BRIDGE – GUPTA BRAND 200662
Optional Acc.
1) Fixed Sine wave oscillator.
2) Sensitive Head Phone/Null Detector
3) Two Dials of 0.1mH & 1mH or Two Dials of 1mH & 10mH
RELIGH BRIDGE – GUPTA BRAND 200661
Optional Acc.
1) Galvalnometer 30-0-30 Division Sensitivity of 2μA/Division
2) Two Dials of 0.1mH & 1mH or Two Dials of 1mH & 10mH or Two
SCHERING BRIDGE FOR CAPACITY MEASUREMENT – GUPTA BRAND 200652
Optional Acc:
- Fixed Frequency Sine Wave Oscillator 1kHz
- Decade Capacitance Box
- Sensitive Head Phone/Null Detector